Fast density reconstruction of Li-BES signal on the COMPASS tokamak
Publication Name: Review of Scientific Instruments
Publication Date: 2018-11-01
Volume: 89
Issue: 11
Page Range: Unknown
Description:
This article describes a fast and automatic reconstruction of the edge plasma electron density from the radiation of energetic Li atoms of the diagnostic beam on the COMPASS tokamak. Radiation is detected by using a CCD camera and by using an avalanche photo-diode system with a temporal resolution of 20 ms and 2 μs, respectively. Both systems are equipped with a 670.8 nm optical filter which corresponds to the lithium 1s22s1-1s22p1 transition. A theoretical model and a data processing procedure of a raw signal to obtain the density profile are described. The reconstruction algorithm provides the absolutely calibrated electron density profiles together with the measurement error estimated from relatively calibrated light profiles; the implementation is performed in Python. Time demanding operations of the code were optimized to provide reconstruction of a single profile within less than 10 ms which makes the code applicable for processing of a large amount of data. Thanks to this calculation speed, it is possible to reconstruct electron density profiles between two consecutive shots on the COMPASS tokamak with 2 μs time resolution.
Open Access: Yes
DOI: 10.1063/1.5049894